Installation of soft X-ray array diagnostics and its application to tomography reconstruction using synthetic KSTAR X-ray images

Cited 4 time in webofscience Cited 4 time in scopus
  • Hit : 295
  • Download : 0
Four-array system of soft X-ray diagnostics was installed on KSTAR tokamak. Each array has 32 viewing chords of two photo-diode array detectors with spatial resolution of 2 cm. To estimate signals from the soft X-ray radiation power, typical n(e), T-e, and argon impurity line radiation profiles in KSTAR are chosen. The photo-diodes were absolutely calibrated as a function of the incident photon energy in 2-40 keV range with a portable X-ray tube. Two-dimensional T-e image properties by multi-energy method were simulated and visualized with six combinations of beryllium filter sets within the dynamic range of signal ratio.