Development of a Kr-83m source for the calibration of the CENNS-10 liquid argon detector

Cited 1 time in webofscience Cited 0 time in scopus
  • Hit : 266
  • Download : 0
We report on the preparation of and calibration measurements with a Kr-83m source for the CENNS-10 liquid argon detector. Kr-83m atoms generated in the decay of a Rb-83 source were introduced into the detector via injection into the Ar circulation loop. Scintillation light arising from the 9.4 keV and 32.1 keV conversion electrons in the decay of Kr-83m in the detector volume were then observed. This calibration source allows the characterization of the low-energy response of the CENNS-10 detector and is applicable to other low-energy-threshold detectors. The energy resolution of the detector was measured to be 9% at the total Kr-83m decay energy of 41.5 keV. We performed an analysis to separately calibrate the detector using the two conversion electrons at 9.4 keV and 32.1 keV.