Sinogram-based super-resolution in PET

Cited 12 time in webofscience Cited 0 time in scopus
  • Hit : 351
  • Download : 0
Spatial resolution is intrinsically limited in positron emission tomography (PET) systems, mainly due to the crystal width. To increase the spatial resolution for a given crystal width, mechanical movements such as wobble and dichotomic motions are introduced to the PET systems. However, multiple sinograms obtained through such movements provide oversampled data. In this paper, to increase the spatial resolution, we present a novel super-resolution (SR) scheme that employs multiple sinograms. For SR, we first propose a blur kernel estimation scheme through a Monte Carlo simulation. Based on the estimated blur kernel, we adopt a maximum a posteriori expectation maximization method in estimating a high-resolution sinogram from multiple low-resolution sinograms. The proposed algorithm provides noticeable improvement of the spatial resolution in real PET images.
Publisher
IOP PUBLISHING LTD
Issue Date
2011-08
Language
English
Article Type
Article
Citation

PHYSICS IN MEDICINE AND BIOLOGY, v.56, no.15, pp.4881 - 4894

ISSN
0031-9155
URI
http://hdl.handle.net/10203/98827