APPLICATION OF A DUAL-ENERGY MONOCHROMATIC X-RAY CT ALGORITHM TO POLYCHROMATIC X-RAY CT: A FEASIBILITY STUDY

Cited 6 time in webofscience Cited 0 time in scopus
  • Hit : 480
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorChang, Sko
dc.contributor.authorLee, HKko
dc.contributor.authorCho, Gyuseongko
dc.date.accessioned2013-03-11T08:16:47Z-
dc.date.available2013-03-11T08:16:47Z-
dc.date.created2012-04-30-
dc.date.created2012-04-30-
dc.date.issued2012-02-
dc.identifier.citationNUCLEAR ENGINEERING AND TECHNOLOGY, v.44, no.1, pp.61 - 70-
dc.identifier.issn1738-5733-
dc.identifier.urihttp://hdl.handle.net/10203/98755-
dc.description.abstractIn this study, a simple post-reconstruction dual-energy computed tomography (CT) method is proposed. A dual-energy CT algorithm for monochromatic x-rays was adopted and applied to the dual-energy CT of polychromatic x-rays by assigning a representative mono-energy. The accuracy of algorithm implementation was tested with mathematical phantoms. To test the sensitivity of this algorithm to the inaccuracy of representative energy value in energy values, a simulation study was performed with mathematical phantom. To represent a polychromatic x-ray energy spectrum with a single-energy, mean energy and equivalent energy were used, and the results were compared. The feasibility of the proposed method was experimentally tested with two different micro-CTs and a test phantom made of polymethyl methacrylate (PMMA), water, and graphite. The dual-energy calculations were carried out with CT images of all possible energy pairs among 40, 50, 60, 70, and 80 kVp. The effective atomic number and the electron density values obtained from the proposed method were compared with theoretical values. The results showed that, except the errors in the effective atomic number of graphite, most of the errors were less than 10 % for both CT scanners, and for the combination of 60 kVp and 70 kVp, errors less than 6.0 % could be achieved with a Polaris 90 CT. The proposed method shows simplicity of calibration, demonstrating its practicality and feasibility for use with a general polychromatic CT.-
dc.languageEnglish-
dc.publisherKOREAN NUCLEAR SOC-
dc.subjectCOMPUTED-TOMOGRAPHY-
dc.subjectATTENUATION CORRECTION-
dc.subjectQUANTITATIVE CT-
dc.subjectATOMIC-NUMBER-
dc.subjectDENSITY-
dc.subjectSPECTRA-
dc.subjectIMPACT-
dc.titleAPPLICATION OF A DUAL-ENERGY MONOCHROMATIC X-RAY CT ALGORITHM TO POLYCHROMATIC X-RAY CT: A FEASIBILITY STUDY-
dc.typeArticle-
dc.identifier.wosid000301275900007-
dc.identifier.scopusid2-s2.0-84863274582-
dc.type.rimsART-
dc.citation.volume44-
dc.citation.issue1-
dc.citation.beginningpage61-
dc.citation.endingpage70-
dc.citation.publicationnameNUCLEAR ENGINEERING AND TECHNOLOGY-
dc.identifier.doi10.5516/NET.08.2010.044-
dc.contributor.localauthorCho, Gyuseong-
dc.contributor.nonIdAuthorLee, HK-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorComputed Tomography(CT)-
dc.subject.keywordAuthorDual Energy-
dc.subject.keywordAuthorMicro CT-
dc.subject.keywordPlusCOMPUTED-TOMOGRAPHY-
dc.subject.keywordPlusATTENUATION CORRECTION-
dc.subject.keywordPlusQUANTITATIVE CT-
dc.subject.keywordPlusATOMIC-NUMBER-
dc.subject.keywordPlusDENSITY-
dc.subject.keywordPlusSPECTRA-
dc.subject.keywordPlusIMPACT-
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 6 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0