Accuracy improvement in the source-based calibration of radiative heat flux sensors by consideration of the temperature non-uniformity of the high temperature blackbody

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 325
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorPark, CWko
dc.contributor.authorLee, DHko
dc.contributor.authorKim, BHko
dc.contributor.authorPark, SNko
dc.contributor.authorShin, Hyun Dongko
dc.date.accessioned2013-03-06T21:28:36Z-
dc.date.available2013-03-06T21:28:36Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2006-02-
dc.identifier.citationMETROLOGIA, v.43, no.1, pp.135 - 141-
dc.identifier.issn0026-1394-
dc.identifier.urihttp://hdl.handle.net/10203/88523-
dc.description.abstractWe have improved the accuracy in source-based calibration of radiative heat flux sensors by considering the temperature non-uniformity of the blackbody cavity. The method measures the responsivity of a heat flux sensor as a function of distance from the blackbody aperture. From this variation of the responsivity, the temperature distribution and the effective emissivity of the blackbody can be determined via a Monte Carlo simulation. The calibration uncertainty is evaluated to be 2.3% (k = 2) for irradiance values up to 10 kW m(-2) at a blackbody temperature around 2900 degrees C. In order to verify the accuracy improvement, the results are compared with those of a detector-based calibration, which demonstrated an agreement within the uncertainty.-
dc.languageEnglish-
dc.publisherIOP PUBLISHING LTD-
dc.subjectMONTE-CARLO METHOD-
dc.subjectEFFECTIVE EMISSIVITIES-
dc.subjectFACILITIES-
dc.subjectCAVITIES-
dc.titleAccuracy improvement in the source-based calibration of radiative heat flux sensors by consideration of the temperature non-uniformity of the high temperature blackbody-
dc.typeArticle-
dc.identifier.wosid000236136000021-
dc.identifier.scopusid2-s2.0-31544460711-
dc.type.rimsART-
dc.citation.volume43-
dc.citation.issue1-
dc.citation.beginningpage135-
dc.citation.endingpage141-
dc.citation.publicationnameMETROLOGIA-
dc.identifier.doi10.1088/0026-1394/43/1/019-
dc.contributor.localauthorShin, Hyun Dong-
dc.contributor.nonIdAuthorPark, CW-
dc.contributor.nonIdAuthorLee, DH-
dc.contributor.nonIdAuthorKim, BH-
dc.contributor.nonIdAuthorPark, SN-
dc.type.journalArticleArticle-
dc.subject.keywordPlusMONTE-CARLO METHOD-
dc.subject.keywordPlusEFFECTIVE EMISSIVITIES-
dc.subject.keywordPlusFACILITIES-
dc.subject.keywordPlusCAVITIES-
Appears in Collection
ME-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike