비등간격 전극열에서의 입자 통과시간을 이용한 전기적 입자속도분포 검출기An Electrical Particle Velocity Profiler Using Particle Transit Time Across Uneven Inter-Gap Electrodes

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 346
  • Download : 0
We present an electrical particle velocity profiler using particle transit time across uneven inter-gap electrodes. We measure both the particle position and velocity from the voltage signals generated by the particles passing across sensing electrodes, thus obtaining the velocity profile of the particles in a microfluidic channel. In the experimental study, we use polystyrene microparticles to characterize the performance of the electrical particle velocity profiler. The particle velocity profile is measured with the uncertainty of 5.44%, which is equivalent to the uncertainty of 5% in the previous optical method. We also experimentally demonstrate the capability of the present method for in-channel clogging detection. Compared to the previous optical methods, the present electrical particle velocity profiler offers the simpler structure, the cheaper cost, and the higher integrability to micro-biofluidic systems
Publisher
대한기계학회
Issue Date
2008-04
Language
Korean
Citation

대한기계학회논문집 A, v.32, no.4, pp.297 - 302

ISSN
1226-4873
URI
http://hdl.handle.net/10203/86554
Appears in Collection
BiS-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0