One micrometer resolution NMR microscopy

Cited 113 time in webofscience Cited 0 time in scopus
  • Hit : 341
  • Download : 0
We obtained a magnetic resonance image of 1 mum resolution and 75 mum(3) voxel volume for a phantom filled with hydrocarbon oil within an hour at 14.1 T. For this work, a specially designed probe with a high sensitivity RF coil and gradient coils generating over 1000 G/cm was built. The optimal pulse sequence was analyzed in consideration of the bandwidth, diffusion coefficients, and T-1 and T-2 relaxations of the medium. The system was applied to the in vivo imaging of a geranium leaf stem to get the images of 2 mum resolution and 200 mum(3) voxel volume. (C) 2001 Academic Press.
Publisher
ACADEMIC PRESS INC
Issue Date
2001-06
Language
English
Article Type
Article
Keywords

MAGNETIC-RESONANCE MICROSCOPY; TEMPERATURE; MICROCOILS; DIFFUSION

Citation

JOURNAL OF MAGNETIC RESONANCE, v.150, no.2, pp.207 - 213