Experimental and theoretical analyses on factors affecting the performance of polypyrrole/glucose oxidase biosensor: Permeability aspect of polypyrrole film

Cited 8 time in webofscience Cited 7 time in scopus
  • Hit : 313
  • Download : 0
The effects of film thickness and enzyme concentration in the electrodeposition solution on the analytical performance of polypyrrole/glucose oxidase (PPy/GOD) biosensor were investigated. The optimum film thickness which maximizes the catalytic current of the electrode was observed at a given enzyme concentration, and this optimum thickness moved down to a thinner region with increasing concentration of enzyme. On the basis of the above results, the conditions for electrochemical deposition of the polypyrrole films were optimized. Theoretical analyses were conducted to elucidate the experimental results. The effect of electrodeposition conditions on the analytical performance of the PPy/GOD biosensor was evaluated in terms of response characteristics, sensitivity, and operational stability.
Publisher
JAPAN SOC ANALYTICAL CHEM
Issue Date
1996
Language
English
Article Type
Article
Keywords

GLUCOSE-OXIDASE; ELECTROCHEMICAL IMMOBILIZATION; ELECTRODE; ENZYME; POLYMERIZATION; THICKNESS

Citation

ANALYTICAL SCIENCES, v.12, no.4, pp.597 - 604

ISSN
0910-6340
DOI
10.2116/analsci.12.597
URI
http://hdl.handle.net/10203/74311
Appears in Collection
BS-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.