Development of single photon tomographic system : GAMMATOM-II單光子 放出形 斷層 촬영기의 開發 : GAMMATOM-II

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 427
  • Download : 0
A high resolution single photon tomographic system for head imaging, GAMMATOMII, was developed. The system consists of 36-NaI(Tl) cylindrical crystals which are arranged in a circular form on the 4 banks, each bank with 9 crystals. The ring artefacts which had appeared in our first system was removed by refining the electronics and mechanical systems. Using the new effective sampling scheme (totally two rotations and one shift motion), sampling of the one eighth of the detector width was successfully achieved. The desired resolution of 7-mm FWHM was achieved with a resolution phantom (20cm diameter) with 99mTc. System design and improvements are reported with the performance characteristics of our system.
Advisors
Cho, Zang-Heeresearcher조장희researcher
Description
한국과학기술원 : 산업전자공학과,
Publisher
한국과학기술원
Issue Date
1982
Identifier
63457/325007 / 000803565
Language
eng
Description

학위논문(석사) - 한국과학기술원 : 산업전자공학과, 1982.2, [ iii, 86 p. ]