Development and performance study of single photon emission computed tomography system : GAMMATOM-Ⅱ단광자 방출형 전산화 단층 촬영기의 개발 및 그의 성능 연구

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 468
  • Download : 0
Single photon emission computed tomography system for the measurement of the distribution of externally administered radionuclides in human body has been developed and its performance is studied. GAMMATOM-II consists of 36 NaI(T$\ell$) scintillation detectors of 2 cm (dia.)$\times$3.8 cm (length). The detectors are arranged into four separate banks each of which consists of nine detectors. Each detector bank is arranged in a circular arc shape and is displaced 1/4 detector width each other from center. Using collimators which have converging multiple holes and new sampling scheme, spatial resolution of 7 mm FWHM was obtained with 15 cm diameter resolution phantom. Sensitivity of the system was observed as counts of 100 (counts/sec)/($\mu$Ci/cc) with 28 cm diameter uniform disk phantom filled with $^{99m}$Tc.
Advisors
Cho, Zang-Heeresearcher조장희researcher
Description
한국과학기술원 : 산업전자공학과,
Publisher
한국과학기술원
Issue Date
1982
Identifier
63453/325007 / 000803553
Language
eng
Description

학위논문(석사) - 한국과학기술원 : 산업전자공학과, 1982.2, [ [ii], 51 p. ]