Complex permittivity extraction of blood glucose at microwave frequencies전자파를 이용한 혈당의 전기적 특성 추출에 대한 연구

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 412
  • Download : 0
In this paper, coaxial sample holder design and its de-embedding and parameter extraction procedure is proposed. The s-parameters were measured up to 1GHz using network analyzer HP8753D and N-type connector, and the de-embedding of N-type connector was performed. We applied proposed de-embedding procedure to extract electrical parameter of blood glucose. As a result, permittivity of blood glucose is extracted. We also analyzed the errors of extracted parameters, which come from instrument error, geometrical error, and temperature variation. Using these error analyses, we reduce the error factor of extracted parameter. We extracted electrical parameters of glucose samples through these all extraction procedure and confirmed the possibility of glucose measurement system based on microwave system.
Advisors
Kim, Joung-Horesearcher김정호researcher
Description
한국과학기술원 : 전기및전자공학전공,
Publisher
한국과학기술원
Issue Date
2001
Identifier
165948/325007 / 000993497
Language
eng
Description

학위논문(석사) - 한국과학기술원 : 전기및전자공학전공, 2001.2, [ vi, 65 p. ]

Keywords

permittivity; glucose; microwave; 유전율; 혈당; 전자팔