Low EMF Design of Cochlear Implant Wireless Power Transfer System using A Shielding Coil

Cited 7 time in webofscience Cited 0 time in scopus
  • Hit : 50
  • Download : 0
Wireless power transfer (WPT) technology is widely used for various applications because of convenience and safety. Especially, medical implant devices such as a cochlear implant are one of the typical application of the WPT technology. However, WPT systems have problems with electromagnetic field (EMF) leakage, which can cause electro-magnetic interference (EMI) issues in the human body. In this paper, we propose an additional shielding coil for the effective suppression of EMF radiation of a WPT system. The proposed method reduces the EMF leakage in operating frequency range. We verified that the proposed shielding coil reduced the EMF leakage by 5.1 dB using 3D EM simulation. Also, the power transfer efficiency (PTE) and link gain stability of the proposed cochlear implant WPT system are designed to work well with proposed shielding coil.
Publisher
Institute of Electrical and Electronics Engineers Inc.
Issue Date
2020-07
Language
English
Citation

2020 IEEE International Symposium on Electromagnetic Compatibility and Signal/Power Integrity, EMCSI 2020, pp.623 - 625

DOI
10.1109/EMCSI38923.2020.9191460
URI
http://hdl.handle.net/10203/311172
Appears in Collection
EE-Conference Papers(학술회의논문)
Files in This Item
There are no files associated with this item.