Study of timing performance parameters for a SiPM-based digital positron annihilation lifetime spectrometer

Cited 1 time in webofscience Cited 0 time in scopus
  • Hit : 112
  • Download : 0
Positron annihilation lifetime spectroscopy (PALS) is a popular method for studying defects in materials. With the development of fast digital oscilloscopes and digitizers, the conventional analog positron annihilation lifetime (PAL) spectrometer is being replaced by a digital PAL spectrometer. Recently, newly conducted research has developed a PAL spectrometer using a silicon photomultiplier (SiPM) instead of a photomultiplier tube (PMT). Timing resolution of PAL spectrometer needs to be improved to identify the lifetimes of positrons. Many parameters affect the timing performance of a SiPM-based digital PAL spectrometer: bias voltage, pulse smoothing, and the constant fraction value of digital constant fraction discrimination (dCFD). Timing resolution depends on the radiation energy, but only a few studies have considered the energy difference of 1274 keV and 511 keV. In this study, we optimized the timing performance parameters of the SiPM-based digital PAL spectrometer by conducting the timing resolution measurements using two γ-ray sources (22Na and 60Co). With a 32 V bias voltage, 5th degree polynomial fitting method, and optimal constant fraction value, we optimized the timing resolution of our SiPM-based digital PAL spectrometer. Furthermore, the optimal constant fraction value depended on the γ-ray energy and we verified it by timing resolution measurement using 207Bi. We studied the reason why performance parameters have optimal conditions and the optimal constant fraction value varies depending on the γ-ray energy. With optimized performance parameters, we measured the lifetime spectrum of pure silicon to confirm the SiPM-based digital PAL spectrometer. The timing resolution function of the SiPM-based digital PAL spectrometer was 185.6 ps. © 2022 IOP Publishing Ltd and Sissa Medialab.
Publisher
IOP Publishing
Issue Date
2022-12
Language
English
Article Type
Article
Citation

JOURNAL OF INSTRUMENTATION, v.17, no.12

ISSN
1748-0221
DOI
10.1088/1748-0221/17/12/C12007
URI
http://hdl.handle.net/10203/303724
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 1 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0