A 67-pJ/bit 435-MHz 16-QAM modulator for capsule endoscopy system with 18-ns start-up using transient DC error correction

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 102
  • Download : 0
We present a 67-pJ/bit 435-MHz modulator for 16 quadrature amplitude modulation with a fast and low-energy start-up for ultra-low-power wireless medical capsule endoscopy application. For a fast start-up, an auto-calibrated transient DC error correction is proposed to minimize the envelope error at the modulator output node occurring right after the initiation of the start-up process. Once the DC error correction is optimized through the calibration procedure, start-up energy of 50 pJ and start-up time of 18 ns are achieved, which enable symbol-level duty cycling. Implemented in a 180-nm CMOS process, the modulator performs 40-Mbps data transmission with -5-dBm output power while consuming 2.66 mW from a 1-V supply. © 2021 IEEE
Publisher
Institute of Electrical and Electronics Engineers Inc.
Issue Date
2021-05
Language
English
Citation

53rd IEEE International Symposium on Circuits and Systems, ISCAS 2021

ISSN
2158-1525
DOI
10.1109/ISCAS51556.2021.9401789
URI
http://hdl.handle.net/10203/288445
Appears in Collection
EE-Conference Papers(학술회의논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0