Performance evaluation of long differential mobility analyzer (LDMA) in measurements of nanoparticles

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 473
  • Download : 528
DC FieldValueLanguage
dc.contributor.authorSong, Dong Keun-
dc.contributor.authorLee, Hye Moon-
dc.contributor.authorChang, Hyuk Sang-
dc.contributor.authorKim, Sang Soo-
dc.contributor.authorManabu, Shimada-
dc.contributor.authorKikuo, Okuyama-
dc.date.accessioned2011-10-25T04:50:06Z-
dc.date.available2011-10-25T04:50:06Z-
dc.date.issued2005-07-
dc.identifier.citation한국입자에어로졸학회 학술대회 논문집, pp.95-96en
dc.identifier.urihttp://hdl.handle.net/10203/25464-
dc.description.abstractPerformance of a long differential mobility analyzer (LDMA) IN Measurements of nanoparticles was evaluated experimentally and numerically. In the evaluation of the LDMA measurements, silver particles in a size range of 5-30 nm were used under increases flow rate. The numerical calculation method was used for calculating the particle trajectory in the LDMA, and the results were used in a comparison with Stolzenburg's transfer function.en
dc.language.isokoen
dc.publisher한국입자에어로졸학회en
dc.subjectLong Differential Mobility Analyzeren
dc.subjectNanoparticleen
dc.titlePerformance evaluation of long differential mobility analyzer (LDMA) in measurements of nanoparticlesen
dc.typeArticleen

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0