Characterization and commissioning of a miniature superficial electronic brachytherapy device for skin cancer treatment피부암 치료를 위한 소형 피상 전자 근접치료장비의 특성평가 및 시운전

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 464
  • Download : 0
In this thesis, the characteristics of the practical application of the superficial electronic brachytherapy device developed for the treatment of skin cancer were found and measured. The measured characteristics are uniformity, absorbed dose rate at skin cancer site. The size of the fabricated surface applicator was 10 mm, 20 mm to be applicable for various skin cancer sizes. First, uniformity was measured. The uniformity was divided into flatness, symmetry, and penumbra. These values were measured and the measured values were verified to have met standards. Second, absorbed dose rate at skin cancer site was measured. In order to measure that, absorbed dose rate at skin surface and percentage depth dose were obtained. After measuring the characteristics, cell experiment and animal experiment were performed to verify the therapeutic effect. Cell experiments and animal experiments have shown that Superficial EBT devices can be used for practical treatment of skin cancer. In addition, similar results with currently used LINAC equipment will be helpful for future commercialization.
Advisors
Cho, Sung Ohresearcher조성오researcher
Description
한국과학기술원 :원자력및양자공학과,
Publisher
한국과학기술원
Issue Date
2017
Identifier
325007
Language
eng
Description

학위논문(석사) - 한국과학기술원 : 원자력및양자공학과, 2017.8,[v, 38 p. :]

Keywords

Superficial Electronic Brachytherapy Device▼aMiniature X-ray Tube▼aSkin Cancer▼aSurface Applicator▼aAnimal Experiment; 소형 피상 전자 근접치료장비▼a소형 X-선 튜브▼a피부암▼a표면 어플리케이터▼a동물실험

URI
http://hdl.handle.net/10203/243207
Link
http://library.kaist.ac.kr/search/detail/view.do?bibCtrlNo=718669&flag=dissertation
Appears in Collection
NE-Theses_Master(석사논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0