Investigation on Beam-Blocker-Based Scatter Correction Method for Improving CT Number Accuracy

Cited 3 time in webofscience Cited 0 time in scopus
  • Hit : 519
  • Download : 0
Cone-beam computed tomography (CBCT) is gaining widespread use in various medical and industrial applications but suffers from substantially larger amount of scatter than that in the conventional diagnostic CT resulting in relatively poor image quality. Various methods that can reduce and/or correct for the scatter in the CBCT have therefore been developed. Scatter correction method that uses a beam-blocker has been considered a direct measurement-based approach providing accurate scatter estimation from the data in the shadows of the beam-blocker. To the best of our knowledge, there has been no record reporting the significance of the scatter from the beam-blocker itself in such correction methods. In this paper, we identified the scatter from the beam-blocker that is detected in the object-free projection data investigated its influence on the image accuracy of CBCT reconstructed images, and developed a scatter correction scheme that takes care of this scatter as well as the scatter from the scanned object.
Publisher
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Issue Date
2017-03
Language
English
Article Type
Article
Keywords

X-RAY SCATTER; COMPUTED-TOMOGRAPHY; IMAGE QUALITY; ALGORITHM; REDUCTION

Citation

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, v.64, no.3, pp.908 - 914

ISSN
0018-9499
DOI
10.1109/TNS.2017.2669519
URI
http://hdl.handle.net/10203/223705
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 3 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0