CALCULATION OF GAMMA SPECTRA IN A PLASTIC SCINTILLATOR FOR ENERGY CALIBRATION AND DOSE COMPUTATION

Cited 3 time in webofscience Cited 0 time in scopus
  • Hit : 523
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorKim, Chankyu-
dc.contributor.authorYoo, Hyunjun-
dc.contributor.authorKim, Yewon-
dc.contributor.authorMoon, Myungkook-
dc.contributor.authorKim, Jong Yul-
dc.contributor.authorKang, Dong Uk-
dc.contributor.authorLee, Daehee-
dc.contributor.authorKim, Myung Soo-
dc.contributor.authorCho, Minsik-
dc.contributor.authorLee, Eunjoong-
dc.contributor.authorCho, Gyuseong-
dc.date.accessioned2017-03-31T05:42:28Z-
dc.date.available2017-03-31T05:42:28Z-
dc.date.created2017-03-20-
dc.date.issued2016-09-
dc.identifier.citationRADIATION PROTECTION DOSIMETRY, v.170, no.1-4, pp.377 - 381-
dc.identifier.issn0144-8420-
dc.identifier.urihttp://hdl.handle.net/10203/222841-
dc.description.abstractPlastic scintillation detectors have practical advantages in the field of dosimetry. Energy calibration of measured gamma spectra is important for dose computation, but it is not simple in the plastic scintillators because of their different characteristics and a finite resolution. In this study, the gamma spectra in a polystyrene scintillator were calculated for the energy calibration and dose computation. Based on the relationship between the energy resolution and estimated energy broadening effect in the calculated spectra, the gamma spectra were simply calculated without many iterations. The calculated spectra were in agreement with the calculation by an existing method and measurements.-
dc.languageEnglish-
dc.publisherOXFORD UNIV PRESS-
dc.subjectRAYS-
dc.titleCALCULATION OF GAMMA SPECTRA IN A PLASTIC SCINTILLATOR FOR ENERGY CALIBRATION AND DOSE COMPUTATION-
dc.typeArticle-
dc.identifier.wosid000384713300085-
dc.identifier.scopusid2-s2.0-84992511249-
dc.type.rimsART-
dc.citation.volume170-
dc.citation.issue1-4-
dc.citation.beginningpage377-
dc.citation.endingpage381-
dc.citation.publicationnameRADIATION PROTECTION DOSIMETRY-
dc.identifier.doi10.1093/rpd/ncw086-
dc.contributor.localauthorKim, Chankyu-
dc.contributor.localauthorYoo, Hyunjun-
dc.contributor.localauthorKim, Yewon-
dc.contributor.localauthorKang, Dong Uk-
dc.contributor.localauthorLee, Daehee-
dc.contributor.localauthorKim, Myung Soo-
dc.contributor.localauthorCho, Minsik-
dc.contributor.localauthorLee, Eunjoong-
dc.contributor.localauthorCho, Gyuseong-
dc.contributor.nonIdAuthorMoon, Myungkook-
dc.contributor.nonIdAuthorKim, Jong Yul-
dc.type.journalArticleArticle-
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 3 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0