Experimental approach to the validity of the hemodynamics extraction method for high resolution near-infrared spectroscopy고해상도 근적외선 영상 장치 (NIRS)를 위한 헤모다이나믹스 추출 방법의 유효성에 대한 실험적 접근

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 591
  • Download : 0
DC FieldValueLanguage
dc.contributor.advisorBae, Hyeon-Min-
dc.contributor.advisor배현민-
dc.contributor.authorChoi, Min-Gyu-
dc.contributor.author최민규-
dc.date.accessioned2013-09-12T02:01:10Z-
dc.date.available2013-09-12T02:01:10Z-
dc.date.issued2013-
dc.identifier.urihttp://library.kaist.ac.kr/search/detail/view.do?bibCtrlNo=513374&flag=dissertation-
dc.identifier.urihttp://hdl.handle.net/10203/180974-
dc.description학위논문(석사) - 한국과학기술원 : 전기및전자공학과, 2013.2, [ ii, 39 p. ]-
dc.description.abstractFor decades, the near-infrared spectroscopy (NIRS) is widely used for a method of non-invasive monitoring to brain. Despite the NIRS have advantages in the feasibility of conversion as portable device and good temporal resolution than other brain reading modalities, its poor spatial resolution of a few centimeters is a critical disadvantage, but it is not easy to be overcome due to the physical characteristics of the NIRS. Given the HEM could improve a spatial resolution of the NIRS as 1cm, but a validity of it was not verified perfectly. Therefore, trying to determine it by experimental approaches in this research. Prior to the experiments because an implementation of adequate system of HEM is considered more im-portant task, I implemented the NIRS system composed of Printed Circuit Board (PCB), headset and S/W. This system satisfied targeted specification and desired arrangement of source-detector could be possible by removable custom-made optode module. Trying to determine the validity of the HEM by a quantitative comparison of between results of conven-tional measure method and the HEM using proposed source-detector structure, but it is improper approach due to time-varying result. In conclusion, I comforted that the scalp effect is a dominant obstacle of NIRS measurement by various experiments and that the assumption of the HEM is valid after a cancellation of the scalp effect.eng
dc.languageeng-
dc.publisher한국과학기술원-
dc.subjectNIRS-
dc.subjectHigh Resolution-
dc.subjectHEM-
dc.subjectHexagon-
dc.subject근적외선 영상장치-
dc.subject고해상도-
dc.subject육각형 구조-
dc.subject헤모다이나믹스 추출 방법-
dc.subjectMIMO-
dc.titleExperimental approach to the validity of the hemodynamics extraction method for high resolution near-infrared spectroscopy-
dc.title.alternative고해상도 근적외선 영상 장치 (NIRS)를 위한 헤모다이나믹스 추출 방법의 유효성에 대한 실험적 접근-
dc.typeThesis(Master)-
dc.identifier.CNRN513374/325007 -
dc.description.department한국과학기술원 : 전기및전자공학과, -
dc.identifier.uid020113636-
dc.contributor.localauthorBae, Hyeon-Min-
dc.contributor.localauthor배현민-
Appears in Collection
EE-Theses_Master(석사논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley