A high-speed miniature screening gaschromatograph with flame ionization detector

Cited 4 time in webofscience Cited 4 time in scopus
  • Hit : 465
  • Download : 3
DC FieldValueLanguage
dc.contributor.authorBanik, Rko
dc.contributor.authorLee, DYko
dc.contributor.authorGweon, Dae-Gabko
dc.date.accessioned2010-02-18T08:02:57Z-
dc.date.available2010-02-18T08:02:57Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2005-12-
dc.identifier.citationJOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, v.19, no.12, pp.2197 - 2204-
dc.identifier.issn1738-494X-
dc.identifier.urihttp://hdl.handle.net/10203/16693-
dc.description.abstractThe combination of Gas chromatography (GC) for separation and Flame Ionization Detection (FID) for detection and identification of the components of a mixture of compounds is a fast and strongly proved method of analytic chemistry. The objective of this research was to design a combined High-speed miniature screening Gas chromatograph along with a Flame Ionization Detector for quick, quantitative and qualitative analysis of gas components. This combined GC-FID system is suitable to detect the volatile and semi-volatile hydrocarbons present in a gas mixture. The construction made it less expensive, easy to use and movable. The complete gas path was developed. On/off valves, temperature and flow sensors and their interface electronics were used for controlling purpose. A Microcontroller was programmed to measure the temperature and gas flow using the sensors and to control and regulate them using the electronics and valves. A pocket PC with its touch screen served as a user interface for the system. Software was developed for the pocket PC, which makes the communication possible with the Microcontroller. The system parameters can be indicated in the Pocket PC as simple text and also the analysis result can be displayed.-
dc.languageEnglish-
dc.language.isoen_USen
dc.publisherKOREAN SOC MECHANICAL ENGINEERS-
dc.titleA high-speed miniature screening gaschromatograph with flame ionization detector-
dc.typeArticle-
dc.identifier.wosid000234000700003-
dc.identifier.scopusid2-s2.0-33644536062-
dc.type.rimsART-
dc.citation.volume19-
dc.citation.issue12-
dc.citation.beginningpage2197-
dc.citation.endingpage2204-
dc.citation.publicationnameJOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY-
dc.identifier.doi10.1007/BF02916459-
dc.embargo.liftdate9999-12-31-
dc.embargo.terms9999-12-31-
dc.contributor.localauthorGweon, Dae-Gab-
dc.contributor.nonIdAuthorBanik, R-
dc.contributor.nonIdAuthorLee, DY-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorgaschromatograph-
dc.subject.keywordAuthorFlame Ionization Detector-
dc.subject.keywordAuthormicrocontroller-
dc.subject.keywordAuthorpocket PC-