A new nano-accuracy AFM system for minimizing Abbe errors and the evaluation of its measuring uncertainty

Cited 33 time in webofscience Cited 43 time in scopus
  • Hit : 428
  • Download : 112
DC FieldValueLanguage
dc.contributor.authorKim, Dongminko
dc.contributor.authorLee, Dong Yeonko
dc.contributor.authorGweon, Dae-Gabko
dc.date.accessioned2010-02-10T06:14:03Z-
dc.date.available2010-02-10T06:14:03Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2007-04-
dc.identifier.citationULTRAMICROSCOPY, v.107, no.4-5, pp.322 - 328-
dc.identifier.issn0304-3991-
dc.identifier.urihttp://hdl.handle.net/10203/16593-
dc.description.abstractA new AFM system was designed for the establishment of a standard technique of nano-length measurement in a 2D plane. In a long range (about several tens of micrometers), measurement uncertainty is dominantly affected by the Abbe error of the XY scanning stage. No linear stage is perfectly straight; in other words, every scanning stage is subject to tilting, pitch and yaw motions. In this paper, an AFM system with minimum offsets of XY sensing is designed. Moreover, the XY scanning stage is designed to minimize the rotation angle, as Abbe errors occur through multiple combination of the offset and the rotation angle. To minimize the rotation angle, an optimal design is performed by maximizing the ratio of the stiffness of the parasitic direction to the motion direction of each stage. This paper describes a design scheme of a full AFM system, in particular, the XY scanner. The full range of a fabricated XY scanner is 100 put x 100 pm. The tilting, pitch and yaw motions are measured by an autocollimator to evaluate the performance of the XY stage. The results show that the AY scanner have a 0.75 arcsec parasitic rotation about the maximum range, thus the uncertainty in terms of the Abbe errors are very small relative to other standard equipment. Using this AFM system, a 3 mu m pitch specimen was measured. The measurement uncertainty of the total system was evaluated especially about pitch length. For a I D evaluation, Abbe errors are the most dominant factor, and the expanded combined uncertainty of the total system was evaluated especially about pitch length. For a I D evaluation, Abbe errors are the most dominant factor, and the expanded combined uncertainty (k = 2) of system was root(4.13)(2) + (5.07 x 10(-5) x p)(2) (nm). For a 2D evaluation, mirror non-orthogonality and Abbe errors are dominant factors, and expanded combined uncertainty (k = 2) of the system was root(4.13)(2) + (1.228 x 10(-4) x p)(2) in the X direction, and root(6 28)(2) + (1.266 x 10(-4) x p)(2) in the Y direction (the unit is nanometers), where p is the measured length in nm. (c) 2006 Elsevier B.V. All rights reserved.-
dc.languageEnglish-
dc.language.isoen_USen
dc.publisherELSEVIER SCIENCE BV-
dc.subjectATOMIC-FORCE MICROSCOPE-
dc.subjectPITCH MEASUREMENTS-
dc.subjectCALIBRATION-
dc.subjectSTANDARDS-
dc.subjectRANGE-
dc.titleA new nano-accuracy AFM system for minimizing Abbe errors and the evaluation of its measuring uncertainty-
dc.typeArticle-
dc.identifier.wosid000244373700007-
dc.identifier.scopusid2-s2.0-33846387372-
dc.type.rimsART-
dc.citation.volume107-
dc.citation.issue4-5-
dc.citation.beginningpage322-
dc.citation.endingpage328-
dc.citation.publicationnameULTRAMICROSCOPY-
dc.identifier.doi10.1016/j.ultramic.2006.08.008-
dc.embargo.liftdate9999-12-31-
dc.embargo.terms9999-12-31-
dc.contributor.localauthorGweon, Dae-Gab-
dc.contributor.nonIdAuthorKim, Dongmin-
dc.contributor.nonIdAuthorLee, Dong Yeon-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorAFM-
dc.subject.keywordAuthorflexure hinge stage-
dc.subject.keywordAuthorabbe error minimization-
dc.subject.keywordAuthorpitch length-
dc.subject.keywordAuthoruncertainty evaluation-
dc.subject.keywordAuthormirror non-orthogonality-
dc.subject.keywordPlusATOMIC-FORCE MICROSCOPE-
dc.subject.keywordPlusPITCH MEASUREMENTS-
dc.subject.keywordPlusCALIBRATION-
dc.subject.keywordPlusSTANDARDS-
dc.subject.keywordPlusRANGE-
Appears in Collection
ME-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 33 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0