Compact soft x-ray transmission microscopy with sub-50 nm spatial resolution

Cited 53 time in webofscience Cited 61 time in scopus
  • Hit : 496
  • Download : 3
In this paper, the development of compact transmission soft x-ray microscopy (XM) with sub-50 nm spatial resolution for biomedical applications is described. The compact transmission soft x-ray microscope operates at lambda = 2.88 nm (430 eV) and is based on a tabletop regenerative x-ray source in combination with a tandem ellipsoidal condenser mirror for sample illumination, an objective micro zone plate and a thinned back-illuminated charge coupled device to record an x-ray image. The new, compact x-ray microscope system requires the fabrication of proper x-ray optical devices in order to obtain high-quality images. For an application-oriented microscope, the alignment procedure is fully automated via computer control through a graphic user interface. In imaging studies using Our compact XM system, a gold mesh image was obtained with 45 nm resolution at x580 magnification and 1 min exposure. Images of a biological sample (Coscinodiscus oculoides) were recorded.
Publisher
IOP PUBLISHING LTD
Issue Date
2006-03
Language
English
Article Type
Article
Citation

PHYSICS IN MEDICINE AND BIOLOGY, v.51, no.6, pp.N99 - N107

ISSN
0031-9155
DOI
10.1088/0031-9155/51/6/N01
URI
http://hdl.handle.net/10203/16579
Appears in Collection
ME-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 53 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley