Surface plasmon resonance-based inhibition assay for real-time detection of Cryptosporidium parvum oocyst

Cited 21 time in webofscience Cited 23 time in scopus
  • Hit : 435
  • Download : 266
DC FieldValueLanguage
dc.contributor.authorKang, CDko
dc.contributor.authorCao, Cko
dc.contributor.authorLee, Jko
dc.contributor.authorChoi, Insungko
dc.contributor.authorKim, BWko
dc.contributor.authorSim, SJko
dc.date.accessioned2009-11-05T01:23:04Z-
dc.date.available2009-11-05T01:23:04Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2008-03-
dc.identifier.citationWATER RESEARCH, v.42, no.6-7, pp.1693 - 1699-
dc.identifier.issn0043-1354-
dc.identifier.urihttp://hdl.handle.net/10203/12123-
dc.description.abstractA surface plasmon resonance (SPR)-based inhibition assay method using a polyclonal antimouse IgM arrayed Cryptosporidium sensor chip was developed for the real-time detection of Cryptosporidium parvum oocysts. The Cryptosporidium sensor chip was fabricated by subsequent immobilization of streptavidin and polyclonal anti-mouse IgM (secondary antibody) onto heterogeneous self-assembled monolayers (SAMs). The assay consisted of the immunoreaction step between monoclonal and-C. parvurn oocyst (primary antibody) and oocysts, followed by the binding step of the unbound primary antibody onto the secondary antibody surface. It enhanced not only the immunoreaction yield of the oocysts by batch reaction but also the accessibility of analytes to the chip surface by antibodyantibody interaction. Furthermore, the use of optimum concentration of the primary antibody maximized its binding response on the chip. An inversely linear calibration curve for the oocyst concentration versus SPR signal was obtained in the range of 1 x 10(6)-1 x 10(2) oocysts ml(-1). The oocyst detection was also successfully achieved in natural water systems. These results indicate that the SPR-based inhibition assay using the Cryptosporidium sensor chip has high application potential for the real-time analysis of C. parvum oocyst in laboratory and field water monitoring. (c) 2007 Elsevier Ltd. All rights reserved.-
dc.languageEnglish-
dc.language.isoen_USen
dc.publisherPERGAMON-ELSEVIER SCIENCE LTD-
dc.subjectSELF-ASSEMBLED MONOLAYERS-
dc.subjectGAD ANTIBODY-
dc.subjectIMMUNOSENSOR-
dc.subjectPROTEIN-
dc.subjectPERFORMANCE-
dc.subjectFABRICATION-
dc.subjectADSORPTION-
dc.subjectBIOSENSOR-
dc.subjectSPR-
dc.titleSurface plasmon resonance-based inhibition assay for real-time detection of Cryptosporidium parvum oocyst-
dc.typeArticle-
dc.identifier.wosid000255166600036-
dc.identifier.scopusid2-s2.0-40749113197-
dc.type.rimsART-
dc.citation.volume42-
dc.citation.issue6-7-
dc.citation.beginningpage1693-
dc.citation.endingpage1699-
dc.citation.publicationnameWATER RESEARCH-
dc.identifier.doi10.1016/j.watres.2007.10.023-
dc.embargo.liftdate9999-12-31-
dc.embargo.terms9999-12-31-
dc.contributor.localauthorChoi, Insung-
dc.contributor.nonIdAuthorKang, CD-
dc.contributor.nonIdAuthorCao, C-
dc.contributor.nonIdAuthorLee, J-
dc.contributor.nonIdAuthorKim, BW-
dc.contributor.nonIdAuthorSim, SJ-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorCryptosporidium parvum-
dc.subject.keywordAuthorinhibition assay-
dc.subject.keywordAuthorreal-time detection-
dc.subject.keywordAuthorsurface plasmon resonance (SPR)-
dc.subject.keywordAuthorself-assembled monolayer (SAM)-
dc.subject.keywordPlusSELF-ASSEMBLED MONOLAYERS-
dc.subject.keywordPlusGAD ANTIBODY-
dc.subject.keywordPlusIMMUNOSENSOR-
dc.subject.keywordPlusPROTEIN-
dc.subject.keywordPlusPERFORMANCE-
dc.subject.keywordPlusFABRICATION-
dc.subject.keywordPlusADSORPTION-
dc.subject.keywordPlusBIOSENSOR-
dc.subject.keywordPlusSPR-
Appears in Collection
CH-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 21 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0