다수의 전기장 분포가 생성되는 단일 미세유로를 이용한 폐암세포 전기천공 및 활성도 분석칩Electroporation and viability monitoring chip for lung cancer cells in single channel with multiple electric field zones

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 506
  • Download : 0
We present an electroporation and viability monitoring chip for lung cancer cells in a single channel with multiple electric field zones. Previous electroporation chips utilized multiple microchannels or electrodes to form multiple electric fields, thus resulting in complex structures. However, the present chip can generate multiple electric fields in a single stepwise microchannel between a pair of electrodes, thus achieving the analysis of both cell electroporation and viability with a simple structure. We demonstrate that the electric field of 0.4 kV/cm results in a maximum percentage of 51.4±3.0% and 26.6±0.7% of viable and electroporated human lung cancer cells, H23 and A549, respectively. The present chip has potential for use in integrated cell chips for transfection studies. © 2012 The Korean Society of Mechanical Engineers.
Publisher
Korean Society of Mechanical Engineers
Issue Date
2012-09
Language
English
Article Type
Article
Citation

TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS, B, v.36, no.9, pp.901 - 905

ISSN
1226-4881
DOI
10.3795/KSME-B.2012.36.9.901