Pinhole collimator design for nuclear survey system

Cited 12 time in webofscience Cited 0 time in scopus
  • Hit : 431
  • Download : 4
A conventional knife-edge collimator, which is widely used in gamma camera for medical diagnosis, is not suitable for nuclear imaging system because many scattering radiations near the pinhole aperture happen and blur image. A new pinhole collimator, which shapes a channeled aperture for reducing image degradation induced by the scattering radiations, is introduced and its characteristics are analyzed by Monte Carlo simulation. Resolutions defined as the full-width at half-maximum (FWHM) of point spread function and efficiencies are calculated about several pinhole diameters from 4 to 8 mm and channel heights from 2 to 10 mm. For this calculation, we assumed that Cs-137 radiation sources with 662 keV mono-energies enter into our designed collimator at the 1 m distance from the detector plane. The efficiencies and resolutions of the channeled collimator are compared with those of the conventional collimator. By comparison results, it is verified that the new collimator takes advantage more than the conventional collimator. The optimum channel height and diameter of the pinhole collimator from simulation results are also proposed and designed. We finally acquired nuclear image mounting this collimator in the nuclear survey system. (C) 2002 Elsevier Science Ltd. All rights reserved.
Publisher
PERGAMON-ELSEVIER SCIENCE LTD
Issue Date
2002-11
Language
English
Article Type
Article
Keywords

IMAGING-SYSTEM

Citation

ANNALS OF NUCLEAR ENERGY, v.29, no.17, pp.2029 - 2040

ISSN
0306-4549
URI
http://hdl.handle.net/10203/10305
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 12 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0