The detective quantum efficiency (DQE) for evaluating the performance of a small gamma camera system with a uniformly redundant array (URA) collimator

Cited 1 time in webofscience Cited 0 time in scopus
  • Hit : 496
  • Download : 142
DC FieldValueLanguage
dc.contributor.authorJeon, Hosangko
dc.contributor.authorCho, Gyuseongko
dc.date.accessioned2009-07-27T04:40:51Z-
dc.date.available2009-07-27T04:40:51Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2008-06-
dc.identifier.citationNUCLEAR INSTRUMENTS METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, v.591, no.1, pp.279 - 281-
dc.identifier.issn0168-9002-
dc.identifier.urihttp://hdl.handle.net/10203/10295-
dc.description.abstractPresently, the gamma scintillation camera is widely used in various industrial, environmental and medical diagnostic fields. Hence, the objective and quantitative evaluation of its imaging performance is required for imaging quality assurance and excessive exposures prevention. In this study, the detective quantum efficiency (DQE) of a small gamma camera with three pinholes (1, 2, and 4 mm in hole diameter) and one-coded aperture (uniformly redundant array (URA); 286 holes, hole diameter 2 mm) collimator was determined using modulation transfer function (MTF), normalized noise power spectrum (NNPS), and incoming signal-to-noise ratio (SNR). We found that the resolution of URA was a little lower than that of the pinhole, 2 mm in hole diameter, and the noise of URA was prominently lower than that of all pinholes. Thus, the DQE of URA was higher than that of the pinhole. We found that the determination of DQE could be a quantitative and effective method to evaluate gamma camera systems performances. (C) 2008 Elsevier B.V. All rights reserved.-
dc.languageEnglish-
dc.language.isoen_USen
dc.publisherELSEVIER SCIENCE BV-
dc.titleThe detective quantum efficiency (DQE) for evaluating the performance of a small gamma camera system with a uniformly redundant array (URA) collimator-
dc.typeArticle-
dc.identifier.wosid000257529300070-
dc.identifier.scopusid2-s2.0-44649103205-
dc.type.rimsART-
dc.citation.volume591-
dc.citation.issue1-
dc.citation.beginningpage279-
dc.citation.endingpage281-
dc.citation.publicationnameNUCLEAR INSTRUMENTS METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT-
dc.embargo.liftdate9999-12-31-
dc.embargo.terms9999-12-31-
dc.contributor.localauthorCho, Gyuseong-
dc.contributor.nonIdAuthorJeon, Hosang-
dc.type.journalArticleArticle; Proceedings Paper-
dc.subject.keywordAuthordetective quantum efficiency (DQE)-
dc.subject.keywordAuthorgamma camera system-
dc.subject.keywordAuthorMTF-
dc.subject.keywordAuthorNNPS-
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 1 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0