DC Field | Value | Language |
---|---|---|
dc.contributor.author | Yun, SD | ko |
dc.contributor.author | Oh, SS | ko |
dc.contributor.author | Han, YJ | ko |
dc.contributor.author | Park, HyunWook | ko |
dc.date.accessioned | 2009-06-30T03:34:08Z | - |
dc.date.available | 2009-06-30T03:34:08Z | - |
dc.date.created | 2012-02-06 | - |
dc.date.created | 2012-02-06 | - |
dc.date.issued | 2009-04 | - |
dc.identifier.citation | JOURNAL OF MAGNETIC RESONANCE IMAGING, v.29, pp.924 - 936 | - |
dc.identifier.issn | 1053-1807 | - |
dc.identifier.uri | http://hdl.handle.net/10203/9897 | - |
dc.description.abstract | Purpose: To develop a novel approach for high-resolution functional MRI (fMRI) using the conventional gradient-echo sequence. Materials and Methods: Echo-planar imaging (EPI) techniques have generally been used for fMRI studies due to their fast imaging time. However, it is difficult for studying brain function at the submillimeter level using this sequence. In addition, EPI techniques have some drawbacks, such as Nyquist ghosts and geometric distortions in the reconstructed images, and subsequently require additional postprocessing to reduce these artifacts. One way of solving these problems is to acquire fMRI data by means of a conventional gradient-echo imaging sequence instead of EPI. To provide a fast imaging time, the proposed method combines higher-order generalized series (HGS) imaging with a parallel imaging technique which is called the HGS-parallel technique. Results: The proposed HGS-parallel technique achieves a 12.8-fold acceleration in imaging time without the cost of spatial resolution. The proposed method was verified through the application of fMRI studies on normal subjects. Conclusion: This study suggests that the proposed method can be used for high-resolution fMRI studies without the geometric distortion and the Nyquist ghost artifacts compared to EPI. | - |
dc.description.sponsorship | Korean Ministry of Health, Welfare and Family Affairs; Contract grant number: 02-PJ3-PG6-EV07-0002 (A020607)Korean Ministry of Education | en |
dc.language | English | - |
dc.language.iso | en_US | en |
dc.publisher | JOHN WILEY & SONS INC | - |
dc.subject | MAGNETIC-RESONANCE | - |
dc.subject | FUNCTIONAL MRI | - |
dc.subject | PHASE CORRECTION | - |
dc.subject | SENSE-EPI | - |
dc.subject | BRAIN | - |
dc.subject | RECONSTRUCTION | - |
dc.subject | ARTIFACTS | - |
dc.subject | ACQUISITION | - |
dc.subject | REDUCTION | - |
dc.subject | REMOVAL | - |
dc.title | High-Resolution fMRI With Higher-Order Generalized Series Imaging and Parallel Imaging Techniques (HGS-Parallel) | - |
dc.type | Article | - |
dc.identifier.wosid | 000264650000024 | - |
dc.identifier.scopusid | 2-s2.0-65249182630 | - |
dc.type.rims | ART | - |
dc.citation.volume | 29 | - |
dc.citation.beginningpage | 924 | - |
dc.citation.endingpage | 936 | - |
dc.citation.publicationname | JOURNAL OF MAGNETIC RESONANCE IMAGING | - |
dc.identifier.doi | 10.1002/jmri.21722 | - |
dc.embargo.liftdate | 9999-12-31 | - |
dc.embargo.terms | 9999-12-31 | - |
dc.contributor.localauthor | Park, HyunWook | - |
dc.type.journalArticle | Article | - |
dc.subject.keywordAuthor | echo-planar imaging (EPI) | - |
dc.subject.keywordAuthor | functional MRI (fMRI) | - |
dc.subject.keywordAuthor | higher-order generalized series imaging | - |
dc.subject.keywordAuthor | high spatial resolution MRI | - |
dc.subject.keywordAuthor | parallel imaging | - |
dc.subject.keywordPlus | MAGNETIC-RESONANCE | - |
dc.subject.keywordPlus | FUNCTIONAL MRI | - |
dc.subject.keywordPlus | PHASE CORRECTION | - |
dc.subject.keywordPlus | SENSE-EPI | - |
dc.subject.keywordPlus | BRAIN | - |
dc.subject.keywordPlus | RECONSTRUCTION | - |
dc.subject.keywordPlus | ARTIFACTS | - |
dc.subject.keywordPlus | ACQUISITION | - |
dc.subject.keywordPlus | REDUCTION | - |
dc.subject.keywordPlus | REMOVAL | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.