Current status of the 40 m small-angle neutron scattering instrument development at the HANARO research reactor

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 870
  • Download : 746
The cold neutron research facility project at the Korean HANARO research reactor was launched in 2003. A state-of-the-art small-angle neutron scattering instrument was selected as a top-priority instrument. An analytical calculation was used to obtain the optimum instrument design and the appropriate range of scattering vectors of the instrument. Then, in order to optimize the instrument layout and to enhance the instrument's performance, a Monte Carlo simulation was performed. The Q range [where Q is the magnitude of the scattering vector given by (4/)sin(/2), is the scattering angle and is the wavelength] of the instrument will extend from 0.0008 to 1.0 Å-1. The instrument layout was optimized based on simulation results. The maximum flux at the sample position can reach about 5.5 × 10 7 n cm-2 s-1.
Publisher
International Union of Crystallography
Issue Date
2007-04-01
Description

*Journal of Applied Crystallography의 Vol. 40, Suppl 1의 경우, Proceedings of the XIIIth International Conference on Small-Angle Scattering (Kyoto, Japan, 9-13, July 2006)과 연계된 저널입니다.

Keywords

small-angle neutron scattering; simulation; Virtual Instrumentation Tool for ESS (VITESS); scattering vector range

Citation

Journal of Applied Crystallography, Vol. 40, Suppl. 1, pp.s442-s446

ISSN
0021-8898
URI
http://hdl.handle.net/10203/11082