Backprojection-filtration reconstruction without invoking a spatially varying weighting factor

Cited 4 time in webofscience Cited 0 time in scopus
  • Hit : 437
  • Download : 0
Methods: The backprojection steps in the existing filtered-backprojection (FBP) and BPF algorithms for fan-beam and cone-beam projections invoke spatially varying weighting factors, which may not only increase the computational load in image reconstruction but also, more importantly, result in reconstruction artifacts. Redundant information in fan-beam projections has been exploited for eliminating the weighting factor in the existing FBP algorithm. However, the new FBP algorithm cannot be applied to image reconstruction in a region of interest from transversely truncated data. In this work, the authors identify approximate data redundancy in circular cone-beam projections and propose a new BPF algorithm in which the approximate data redundancy is exploited for eliminating the spatially varying weighting factor in the existing BPF algorithm. Results: The authors have implemented and evaluated the proposed BPF algorithm in numerical studies of reconstructing 3D images from both the nontruncated and truncated projection data in a circular cone-beam scan. The results of numerical studies demonstrate that the proposed BPF algorithm retains the resolution property of the existing BPF algorithm, and that it can also reconstruct accurately ROI images from truncated data. More importantly, the results also indicate that the proposed BPF algorithm not only is computationally more efficient but also yields generally lower image variances than the existing BPF algorithm. Conclusions: A BPF algorithm was proposed that not only retains the desirable properties of the existing BPF algorithm but also possesses improved computational and noise properties over the latter.
Publisher
AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
Issue Date
2010-03
Language
English
Article Type
Article
Keywords

CONE-BEAM CT; IMAGE-RECONSTRUCTION; FAN-BEAM; COMPUTED-TOMOGRAPHY; ALGORITHM; NOISE; REGIONS; QUALITY; SCAN

Citation

MEDICAL PHYSICS, v.37, no.3, pp.1201 - 1209

ISSN
0094-2405
DOI
10.1118/1.3285041
URI
http://hdl.handle.net/10203/100666
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 4 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0